Ask The ExpertHarry, how much time do I have to fill the Melt Indexer Barrel and start the test in a Method A Test. I have heard 60 Seconds to fill and then wait 300 Seconds or 420 Seconds. What is the proper procedure?
Answer
ASTM requires the operator to fill the bore, insert the piston and start the test within a minute. As soon as it is finished, the test is started on the machine and the preheat countdown begins. Generally, ASTM D1238 requires 420 seconds +/- 30 seconds for the pre-heat period but that can be overruled in the materials specification if there is one.