Harry, how much time do I have to fill the Melt Indexer Barrel and start the test in a Method A Test. I have heard 60 Seconds to fill and then wait 300 Seconds or 420 Seconds. What is the proper procedure?

Author:
Tap Water
Date added:
Wednesday, 22 August 2007
Last revised:
Thursday, 23 August 2007
Hits:
558
Rating:
 
Vote for this:
Good - Bad
favoured:
0 Favour

Answer

ASTM requires the operator to fill the bore, insert the piston and start the test within a minute. As soon as it is finished, the test is started on the machine and the preheat countdown begins. Generally, ASTM D1238 requires 420 seconds +/- 30 seconds for the pre-heat period but that can be overruled in the materials specification if there is one.